Quantitative Analysis of Warfarin Tablets Containing Salt-Form Impurities Using Transmission Raman Spectroscopy
21 May 2019This application note demonstrates the use of transmission Raman spectroscopy (TRS) for the quantification of Warfarin in whole intact pharmaceutical tablets at 0.5% w/w with 20 second measurement time. TRS can differentiate between and quantify two forms of Warfarin without sample preparation. Traditional chromatography techniques cannot do this as the form information is destroyed by solvation. TRS enables fast, automated, non-destructive quantitative bulk analysis of pharmaceutical oral solid dose forms for content uniformity and polymorph analysis.
Links
Tags
Raman SpectroscopyRaman spectroscopy is used to discern the vibrational and rotational states of molecules and hence the chemical composition of a sample by measuring the inelastic scattering of monochromatic light. Explore a range of Raman spectrometers, including handheld/portable Raman spectrometers for QC/QA labs and in situ spectrometers for processes. Conduct Raman imaging for microanalysis of mixed samples using a Raman microscope. Raman spectrographs are also available. Find the best Raman spectroscopy products in our peer-reviewed product directory: compare products, check customer reviews and receive pricing direct from manufacturers.PharmaceuticalsPharmaceuticals are medicinal drugs used in healthcare to diagnose, prevent, cure and treat illnesses. Pharmaceuticals that are excreted after use appear in wastewater and can have detrimental effects on the environment.