Ionic Strength Dependence of the Heating Time in T-Jump Instrument TJ-64
10 Oct 2011This application note shows that the observed rate of heating (kobs) depends upon the ionic strength of the NATA solution. The magnitude of kobs is determined by electronic filtering where longer time constants yield an improved signal-to-noise ratio (SNR) but limit the temporal resolution of the reaction of interest. To measure the fastest reaction rates in the TJ-64, capacitor discharge T-jump instrument, it was necessary to use an ionic strength of at least 300 mM.
Tags
Fluorescence SpectroscopyFluorometers and spectrofluorometers (also called fluorescence spectrometers) are used to measure the intensity and wavelength of fluorescent light emitted from a sample after excitation by illumination. Spectrofluorometers utilize monochromators to select the desired wavelengths, whereas filter fluorometers employ a set of filters. Spectrofluorometers for measuring steady-state fluorescence and lifetime fluorescence (or time-resolved fluorescence) are available, as well as fluorescence microscopes and microplate readers. Find the best fluorescence spectroscopy products in our peer-reviewed product directory: compare products, check customer reviews and receive pricing direct from manufacturers.