ResourceLab Automation
Gravimetric calibration and testing of piston-operated volumetric apparatus
30 Mar 2023Reliable and reproducible liquid handling results depend on the regular testing and calibration of instruments. Globally, the most accepted International Standardization Organization (ISO) standard that details requirements for producing and in-use control of piston-operated volumetric apparatus (POVA) is the ISO 8655. In this white paper, Sartorius presents the requirements for gravimetric calibration and testing of POVA according to part 6 and part 7 of the ISO 8655, revised in 2022. Furthermore, the differences between calibration and testing according to chapters 6 and 7, which describe the testing method, testing environment, testing equipment, and reporting requirements, will be explored.
Links
Tags
Automated Liquid HandlingAutomated liquid handling equipment is used to mix, dilute and dispense allotted volumes of liquid into microplates and other vessels automatically. The robotic, liquid handling arms can dispense from single channel to 3456 multichannel as well as operate nanoliter dispensing, enabling high throughput of samples. Find the best automated liquid handlers in our peer-reviewed product directory: compare products, check customer reviews and receive pricing direct from manufacturers.Lab EssentialsLab essentials include laboratory consumables that are vital to the operation of all laboratories.Liquid HandlingLiquid handling involves the precise transfer, mixing, and dispensing of liquids in laboratory settings. It is essential for accurate and consistent results in biological, chemical, and pharmaceutical research. Automated liquid handling systems improve efficiency and reduce human error in sample preparation and analysis. Explore liquid handling tools in our peer-reviewed product directory; compare products, read customer reviews, and get pricing directly from manufacturers.CalibrationPrecision Liquid Handling