ABX Micros ES 60
Overview Integrated barcode reader USB connection Unique pack system Integrated ticket printer Features RBC/PLT Detection Principles: Method Impedance, Ruby diameter 50 μm, Depression of count 200 mb, Duration of count 2 (or 3) x 6 seconds, Dilution ratio 1/15 000 HGB Measurement: Method Photometry, Wavelength 550 nm, Dilution ratio 1/250 HCT Measurement: Method Numeric integration WBC Measurement: Metho…

The supplier does not provide quotations for this product through SelectScience. You can search for similar products in our Product Directory.
I can recommend this to diagnostics
Serum analysis for complete blood picture
Very good equipment and user friendly software and quick analysis of results with accurate value
Review Date: 25 Mar 2025 | Horiba Medical
Overview
- Integrated barcode reader
- USB connection
- Unique pack system
- Integrated ticket printer
Features
- RBC/PLT Detection Principles: Method Impedance, Ruby diameter 50 μm, Depression of count 200 mb,
- Duration of count 2 (or 3) x 6 seconds, Dilution ratio 1/15 000
- HGB Measurement: Method Photometry, Wavelength 550 nm, Dilution ratio 1/250
- HCT Measurement: Method Numeric integration
- WBC Measurement: Method Impedance, Ruby diameter 80 μm, Depression of count 200 mb, Duration of count 2 (or 3) x 6 seconds, Dilution ratio 1/260
- 3 Part Differentiation: Method Impedance + specific lyse action
- MCV, MCH, MCHC, RDW, PCT*, PDW* Calculation
- ABX Micro-sampling: Only 10μl of whole blood per analysis
- ABX Liquid valves: Precise volume and reliability
- ABX stepper motor vacuum pump: Noiseless running (no-compressor)
* RUO parameters (Research Use Only)